OI, OSI, MAP, HFOV
The Ventilator Parameters Calculator is a multi-tab NICU tool covering conventional ventilation, high-frequency oscillatory ventilation (HFOV), and non-invasive CPAP support. It calculates oxygenation indices used to grade the severity of hypoxaemic respiratory failure and to help decide when to escalate or de-escalate respiratory support. The tool covers Mean Airway Pressure (MAP), the Oxygenation Index (OI) and Oxygen Saturation Index (OSI), HFOV-specific DCO2 (CO2 diffusion) calculations, and CPAP/high-flow weaning indices (SOPI, S/F ratio, ROX index).
OI combines Mean Airway Pressure, FiO2, and PaO2 into a single number used to grade severity of hypoxaemic respiratory failure. The app bands OI as Mild (<5), Moderate (5-15), Severe (16-25), Very Severe (26-40), or a range where ECMO should be considered (>40-60).
OI uses PaO2 from an arterial blood gas, while OSI substitutes pulse-oximetry SpO2 so it can be tracked non-invasively. The tool cross-checks the two using OI approximately equal to (2.3 x OSI) - 4.
The calculator offers Square (K=1.0, the default for most conventional pressure-control/SIMV ventilators), Sine (K=0.64, used for HFOV and some sine-flow volume-controlled modes), and Triangular (K=0.5, decelerating/ramp flow patterns) — pick the one matching the ventilator's actual pressure waveform.
The severity table flags ECMO evaluation when OI exceeds 40 on maximum support (HFOV plus inhaled nitric oxide) for more than 4-6 hours, or at any time OI exceeds 60.
ROX = (SpO2/FiO2) / respiratory rate is presented in the CPAP/non-invasive tab as a weaning predictor, with a value greater than 4.88 at 2 hours associated with high-flow nasal cannula (HFNC) success.